GE TransPort PT878GC User Manual page 118

Portable gas flowmeter
Table of Contents

Advertisement

April 2009
Entering a Calibration Factor (cont.)
1. The first prompt asks if you wish to enable the calibration
factor.
a. Use the [ ] and [ ] keys to scroll to the appropriate
radio button.
b. Press
[ENTER]
2. The next prompt asks if you want a single K factor or a table
of K factors.
a. Use the [ ] and [ ] keys to scroll to the appropriate
radio button.
b. Press
[ENTER]
3. The following steps depend on whether you select a single
value or a table.
If you entered Single:
The prompt asks for a value.
a. Press
[ENTER]
b. Use the numeric keys to enter the recommended value.
Generally, if you have enabled the Reynolds Correction
factor, the correction factor should be set to 1.00.
Otherwise, the typical factor is between 0.5 and 2.00.
c. Press
[ENTER]
You have completed entering data for a single Calibration Factor.
If you entered Table:
The menu asks, first for the data source, and then for the entry of
values into the Correction Factor table.
a. Press
[ENTER]
Measurement window (Figure 4-19 on page 4-40) appears.
4-54
to confirm your selection.
to confirm your selection.
to open the text box.
to confirm your selection.