Entering the Digital Output (cont.)
If you entered Test Points:
The prompt asks for the signal type from two choices: transmit or
receive.
a. Press
[ENTER]
b. Scroll to the required signal.
c. Press
[ENTER]
If you entered Gate Input:
1. The first prompt asks for the gate active.
a. Press
[ENTER]
b. Scroll to the gate, either Contact Open or Contact Closed.
c. Press
[ENTER]
2. The remaining prompt asks for the mode, automatic or
manual.
a. Press
[ENTER]
b. Scroll to the required mode.
c. Press
[ENTER]
Stopwatch Totalizer
Through the Gate Input option, you can implement the
Stopwatch Totalizer function to measure totals manually. To set
up the Totalizer function:
1. At the Gate Active prompt, select Contact Closed.
2. At the Mode prompt, select Manual, and press
The stopwatch icon (see page 2-11) appears in the system tray. To
start or stop the function, press the minus (-) key on the keypad.
You have completed entering parameters in the Digital Output
option.
To confirm the entries and return to Operate mode, press
(OK).
To leave the window without confirming the entries, press
(Cancel) or the
Programming Site Data
to open the drop-down menu.
to confirm your selection.
to open the drop-down menu.
to confirm your selection.
to open the drop-down menu.
to confirm your selection.
[ESC]
key.
April 2009
[ENTER]
.
[F3]
[F2]
4-45