Path Length; Axial Length; Reynolds Correction; Calibration Factor - GE DigitalFlow GF868 Programming Manual

Ultrasonic flowmeter for flare gas (1-channel)
Hide thumbs Also See for DigitalFlow GF868:
Table of Contents

Advertisement

1.8.4 Path Length

= inch or
5. Press
[F1]
. (The meter will only accept values from 1/8 to 900 in.)
[ENT]
Note: The factory has calculated both the transducer signal path length
based on the exact transducer configuration used for the application. These values are engraved on the
flowcell and/or are included in the documentation supplied with the meter.

1.8.5 Axial Length

= inch or
6. Press
[F1]
[ENT]
.
Note: The factory has calculated both the transducer signal path length
based on the exact transducer configuration used for the application. These values are engraved on the
flowcell and/or are included in the documentation supplied with the meter.

1.8.6 Reynolds Correction

to turn Reynolds Correction off, or [F2] to turn it on.
7. Press
[F1]
Note: Reynolds Correction is a number based on the Kinematic Viscosity and flow rate of the fluid. It should be
enabled for most applications.
a. When you enable the Reynolds Correction Factor, you must also enter the Kinematic Viscosity of your gas, as
listed in Sound Speeds and Pipe Size Data. Use the numeric keys to enter a value, and press [ENT].

1.8.7 Calibration Factor

8. Enter a value for the flow Calibration Factor and press
values between 0.50 and 2.00.)
1.8.7a
Procedure Options
After completing the above steps, the meter returns to the
To continue programming the meter, refer to the menu maps in Appendix A and navigate to the desired menu.
Then, proceed to the appropriate section of the manual for instructions.
To leave the User Program and retain the previous settings, press
prompt. Any programming changes will be discarded and you will be returned to the data display.
To leave the User Program and return to measurement mode, press
prompt. Your programming changes will be entered into the meter's memory, and you will be returned to the data
display.
DigitalFlow™ GF868 Programming Manual (1-Channel)
[F2] = feet
to select the units. Then, enter the Path Length (P) of the ultrasonic signal. Press
to select the units.Then, enter the Axial Length (L) of the ultrasonic signal and press
[F2] = feet
(P)
and the transducer signal axial length
and the transducer signal axial length
(P)
. The default value is 1.00. (The meter will only accept
[ENT]
prompt. Continue as follows:
User PROGRAM
and then press
[EXIT]
and then press
[EXIT]
Chapter 1. Programming Site Data
at the SAVE
[F1] = NO
at the SAVE
[F2] = YES
(L)
,
,
(L)
13