8.6.2.2

Installing cables on a type B lifter column - part 2

·
Run the cable bundle, including the protective tube for the cables, through the opening in the
front cover (2).
·
Secure the cable bundle with a cable tie under the cover.
·
Raise and lower the seat several times while closely watching the cable bundle.
The cable bundle must not get trapped, stretched or crushed. The cable bundle must move
sideways, as shown in the following illustrations. The drive motor must not be touched by the cable
bundle.
Lifter column completely extended.
76
Invacare
Lifter column completely retracted.
Pay attention to areas where the cable bundle might
get crushed or touch the motor.
®