8.12.1.3 Proceed with removing the motor
●
Remove the batteries as described in chapters 8.4.1 and 8.4.2.
ATTENTION: Danger of crushing!
The True Track® Plus motor unit is extremely heavy. Injury hazards to hands!
· When disassembling, note that the True Track® Plus motor is extremely heavy!
●
Loosen and remove the four fixation screws (1) of the motor.
●
Remove the shock absorber mounting (7).
●
Remove the brake lever (3) and the threaded rod for the brake lever (5).
●
Remove the motor.
4 ®
Series
SERVICE MANUAL
133