Invacare
- STORM
·
Remove the cable from the electronics mountings (6).
·
Pull the snap fasteners (3) next to the True Track® Plus power module (4) to the rear.
·
Lift the True Track® Plus power module together with the mounting (2) and remove the
mounting from the battery flap (1).
·
Loosen and remove the three countersunk screws (5) with a 3 mm Allen key.
·
Remove the True Track® Plus power modul.
·
Undo the cable cover caps on the motors using a 3 mm Allen key.
·
Undo the ring terminal ends on the motors using a 2.5 mm Allen key.
·
Replace the True Track® Plus power module.
Refitting the True Track
·
Refit the parts in reverse order.
·
Plug all the plugs into their old positions.
·
Refix the cable in the electronics mountings.
·
Push the motor cable terminal ends in from the side and fit using a torque wrench. Tighten the
bolts with 2.5 Nm.
·
Select the drive program as described in chapter 8.2.9.
·
If a new software version is available, update the drive program as described in chapter 8.3.
·
Adapt the drive program using the programming software if necessary.
·
To conclude, check all the mobility aid functions.