Invacare Storm4 Series Service Manual page 209

Hide thumbs Also See for Storm4 Series:
Table of Contents

Advertisement

®
Invacare
- STORM
·
Place the adapter (4) in the seat support (3) from below.
·
Fix the adapter with the two Allen screws (1).
·
Replace the seat support on the lifter pillar (5).
CAUTION!
Damage hazard to lifter pillar! Lifter pillar can be damaged if screws are too long. The
screws used may protrude by a maximum of 8.5 mm from the adapters.
· Only use the screws supplied.
· Check the screw length precisely if you do use other screws.
·
Apply a medium-strength screw blocker system to the four Allen screws (2).
·
Insert the four Allen screws (2) and tighten them to a torque of 25 Nm.
·
Refit the other parts in reverse order.
·
To conclude, check all the mobility aid functions.
4 ®
Series
SERVICE MANUAL
209