Calibrating True Track® Plus Motors - Invacare Storm4 Series Service Manual

Hide thumbs Also See for Storm4 Series:
Table of Contents

Advertisement

SERVICE MANUAL
Remove the temporary
M8 x 30 mm screw at
Position 5: Reposition the
original M8 x 20 mm screw and
tighten by hand (max. 8 Nm).
Remove the temporary
M8 x 30 mm screw at
Position 1: Reposition the
original M8 x 20 mm screw and
tighten by hand (max. 8 Nm).
Remove the temporary
M8 x 30 mm screw at
Position 3: Reposition the
original M8 x 20 mm screw and
tighten by hand (max. 8 Nm).
Tighten all screws to 25 Nm.
8.12.2 Calibrating True Track® Plus motors
The following sections explain the calibration process with the handheld programming device.
WARNING: Risk of accidents!
Hazards to workers, surroundings and mobility device!
· Do not leave the mobility device unattended during the following procedure!
· Make sure that BOTH drive wheels are raised and CANNOT touch the ground.
· Secure the area.
Requirements:
· Dynamic DX HHP" handheld programming device
Prop up the wheelchair with wooden blocks. The drive wheels must not be touching
the floor or the work surface.
Connect the programming device.
The programming device
shows the following:
DX HHp V1.20
Select a language...
GB
D
View or edit
System?
YES
?
Technician mode
Enter Password
000
EXIT
D1
136
You should enter this:
"GB"
NL
S
"TECH"
DIAG TECH
Enter code "592" with keys D1 to D3, then select
"EXIT".
D2
D3
®
Invacare
- STORM
4 ®
Series