SERVICE MANUAL
●
Remove the temporary
M8 x 30 mm screw at
Position 5: Reposition the
original M8 x 20 mm screw and
tighten by hand (max. 8 Nm).
●
Remove the temporary
M8 x 30 mm screw at
Position 1: Reposition the
original M8 x 20 mm screw and
tighten by hand (max. 8 Nm).
●
Remove the temporary
M8 x 30 mm screw at
Position 3: Reposition the
original M8 x 20 mm screw and
tighten by hand (max. 8 Nm).
●
Tighten all screws to 25 Nm.
8.12.2 Calibrating True Track® Plus motors
The following sections explain the calibration process with the handheld programming device.
WARNING: Risk of accidents!
Hazards to workers, surroundings and mobility device!
· Do not leave the mobility device unattended during the following procedure!
· Make sure that BOTH drive wheels are raised and CANNOT touch the ground.
· Secure the area.
Requirements:
· Dynamic DX HHP" handheld programming device
●
Prop up the wheelchair with wooden blocks. The drive wheels must not be touching
the floor or the work surface.
●
Connect the programming device.
The programming device
shows the following:
DX HHp V1.20
Select a language...
GB
D
View or edit
System?
YES
?
Technician mode
Enter Password
000
EXIT
D1
136
You should enter this:
"GB"
NL
S
"TECH"
DIAG TECH
Enter code "592" with keys D1 to D3, then select
"EXIT".