Invacare
- STORM
The programming device
shows the following:
Technician mode
Master JS Module
JOYSTICK CALIBRATION
EXIT YES
** MAIN MENU **
View or edit GB
Power Module ?
NEXT
GB Controller
Torque
XX %
EXIT NEXT
GB Controller
Tremor Damping
XX %
EXIT NEXT
GB Controller
Speed Progression
XX %
EXIT NEXT
GB Controller
Turn Progression
XX %
EXIT NEXT
GB Controller
Calibrate
Motors?
EXIT NEXT
GB MOTOR CALIBRATION
Wheels will move!
Drive wheels raised?
EXIT
GB MOTOR CALIBRATION
Chair will drive!
Are wheels raised?
EXIT
GB MOTOR CALIBRATION
-BEGIN- to start.
Wheels will drive!
EXIT
4 ®
Series
You should enter this:
"NEXT"
NEXT
"YES"
YES
"NEXT"
DOWN
UP
"NEXT"
DOWN
UP
"NEXT"
DOWN
UP
"NEXT"
DOWN
UP
"YES"
YES
"YES" (if drive wheels have been raised)
YES
"YES" (if drive wheels have been raised)
YES
"BEGIN" (if drive wheels have been raised)
BEGIN
SERVICE MANUAL
137