Invacare Storm4 Series Service Manual page 36

Hide thumbs Also See for Storm4 Series:
Table of Contents

Advertisement

SERVICE MANUAL
BLINK CODE
9 or 10 x blink
11 x blink
12 x blink
36
POSSIBLE CAUSE
Faulty data transmission
between modules
Drive motors overloaded /
overheated
Compatibility problems
between modules
Invacare
SOLUTION
Remove electronic modules
except for the power
module and the remote.
Replace the modules one
after another in order to
ensure which was the one
causing the fault.
Switch remote on and off /
wait if necessary
Remove incorrect module
®
4 ®
- STORM
Series
DOCUMENTATION
See chapter 8.2.2.
-
See chapter 8.2.2.