8.17.3 Replacing the servo motor

Requirements:
· 4 mm Allen key
· 5 mm Allen key
· 10 mm socket spanner
· 13 mm jaw spanner
· Size 2 Phillips recessed head spanner
· Snap ring pliers
Fitting the planetary gear:
·
Disconnect the two motor plugs (1).
·
Dismantle the entire back section and remove both planetary gears as described in chapter
8.17.2.
·
Loosen and remove the two nuts (2): only the upper nut is visible in the illustration.
·
The motor (5) and the gear (4) form one unit. Pull them from the axle together (3).
·
Replace the motor/gear unit.
Refitting the planetary gear:
·
Refit the parts in reverse order.
·
To conclude, check all the mobility aid functions.