SERVICE MANUAL
PROBLEM
Electrical
adjustment
motor does
not react
1 Only applies to remote SD24
2 All remotes except SD24
34
OTHER
POSSIBLE
SYMPTOMS
CAUSE
Remote shows
Lighting /
1
blinking "E"
or
actuator module
2
blink code 2
defective
Status diode on
lighting/actuator
module does not
go out even if
the remote has
been switched
off or
disconnected
None
Cable
disconnected or
damaged
Electrical
adjusting motor
defective
Remote
defective
®
Invacare
SOLUTION
Documentation
Replace lighting /
See chapter 8.2.6.
actuator module
Safeguard cable
See chapter 8.6.
connection,
replace cable if
necessary
Check adjusting
motor
Replace remote
See chapter 8.2.1.