Invacare Storm4 Series Service Manual page 57

Hide thumbs Also See for Storm4 Series:
Table of Contents

Advertisement

®
Invacare
- STORM
·
Pull the locking pin (1) that is located next to the power module backwards.
·
Lift the power module up, together with the bracket, and remove the bracket from the battery
compartment cover.
·
Place the power module, together with bracket, on top of the battery box or the actuator
module, if installed.
·
Press both catches (3) inwards and open the battery box cover.
·
Use the 5 mm Allen key to loosen and remove both screws (2).
·
Remove the safety catch (1) and the G-Trac bracket (3).
·
Loosen and remove the self-locking nut (5) including the washer with a 5 mm Allen key and a
10 mm socket wrench.
·
Remove the G-Trac sensor.
·
Disconnect the cable of the G-Trac sensor from the electronics module.
·
Replace the G-Trac sensor.
4 ®
Series
SERVICE MANUAL
57

Advertisement

Table of Contents
loading

Table of Contents