8.11.3 Replacing the servo motor

Requirements:
• 4 mm Allen key
• 5 mm Allen key
• 10 mm socket spanner
• 13 mm jaw spanner
• Size 2 Phillips recessed head spanner
• Snap ring pliers
Fitting the planetary gear:
Disconnedct the two motor plugs (1).
Dismantle the entire back section and remove both planetary gears as described in chapter
Fehler! Verweisquelle konnte nicht gefunden werden..
Loosen and remove the two nuts (2): only the upper nut is visible in the illustration.
The motor (5) and the gear (4) form one unit. Pull them from the axle together (3).
Replace the motor/gear unit.
Refitting the planetary gear:
Refit the parts in reverse order.
To conclude, check all the mobility aid functions.