P805 Echo Confidence
Displays the echo confidence of the measurement echo from the last shot.
Primary Index
Values
Related
Use this feature to monitor the effect of transducer aiming, location, and mechanical
transducer / mounting isolation.
Both short and long shot Echo Confidence is displayed. (To display this value in the
auxiliary display while the unit is running, press
Display
x:--
--:y
x:y
E
--:--
P806 Echo Strength
Displays the strength (in dB above 1 uV RMS) of the echo which was selected as the
measurement echo.
Primary Index
Values
P807 Noise
Displays the average and peak ambient noise (in dB above 1 uV RMS) being processed.
Primary Index
Values
The noise level is a combination of transient acoustic noise and electrical noise (induced
into the transducer cable or receiving circuitry). See
Troubleshooting
Page 202
Transducer
Format: x:y (view only)
x = short (0 to 99)
y = long (0 to 99)
P804 Confidence Threshold
P830 TVT Type
short shot confidence value, (long shot not used).
long shot confidence value, (short shot not used).
short and long shot confidence values (both used).
transducer cable is open or short circuited.
no shots were processed for Sonic Intelligence® evaluation.
Transducer
Format: 0 to 99 (view only)
Transducer
Format: x:y (view only)
x = average (-99 to 99)
y = peak (-99 to 99)
Section on page 232.
HydroRanger 200 – INSTRUCTION MANUAL
for 4 seconds.)
Description
Noise Problems
in the
7ML19985FC03