P654 Velocity at 20°C
This value is used to automatically calculate Sound Velocity (P653).
Primary Index
Values
Related
After performing a Sound Velocity Calibration, check this value to verify the acoustic
beam atmosphere is air (344.1 m/s or 1129 ft/s).
Alternatively, if the acoustic beam atmosphere sound velocity at 20°C (68 °F) is known,
and the sound velocity vs. temperature characteristics are similar to that of air, enter the
sound velocity.
The units used are m/s if P005 = 1, 2, or 3 (or ft/s if P005 = 4 or 5).

Temperature Compensation (P660 to P664)

P660 Temp Source
Source of the temperature reading used to adjust the speed of sound.
Primary Index
Values
Alters
Related
The HydroRanger 200 uses the TS-3 temperature sensor assigned to the transducer. If
one is not connected, the ultrasonic/temperature transducer is used. If the transducer
does not have an internal temperature sensor, the Temp Fixed (P661) value is used.
If the acoustic beam atmosphere temperature varies with distance from the transducer,
connect a TS-3 Temperature Sensor and ultrasonic / temperature transducers, and select
average.
7ML19985FC03
Single Point Model
Global
Range: 50.0 to 2000.0 m/s (164 to 6562 ft/s)
P005 Units
P651 Sound Velocity Calibration
P653 Velocity
Transducer
1
*
AUTO
2
Temp Fixed
3
Ultrasonic/Temperature Transducer
4
TS-3 Temperature Sensor
5
Average (TS-3 and transducer)
P664 Temperature
P651 Sound Velocity
P653 Velocity
P654 Velocity at 20°C
P661 Temp Fixed
HydroRanger 200 – INSTRUCTION MANUAL
Dual Point Model
Transducer
Page 181