Display And Reading (P060 To P062) - Siemens HydroRanger 200 Instruction Manual

Table of Contents

Advertisement

Entering a Level Breakpoint
1.
Go to Parameter P054.
2.
For each index enter a breakpoint in measurement units.
3.
Ensure that each breakpoint corresponds to the same index for P055.
P055 Volume Breakpoints and Characterization
(Universal Volume Calculation) [6 relay model]
Each segment defined by the level breakpoints (P055) requires a volume so that the
HydroRanger 200 can make the level-to-volume calculations.
Primary Index
Secondary Index
Values
Related
Typical volume calculations
Cone
2
V = (1/3)πr
h
Entering a Volume Breakpoint
1.
Go to Parameter P055.
2.
For each index enter a volume.
3.
Ensure that each volume corresponds to the same index for P054.
Volume Characterization
For more on

Display and Reading (P060 to P062)

These parameters are used to:
Change the number of decimal places displayed
Convert the Reading to alternate units
Reference measurements to other than Empty (P006) or Span (P007)
Page 126
Level
Breakpoint
Range: 0.0 to 99999
P054 Breakpoints Levels (Universal Volume Calculation)
Cylinder
2
V = πr
h
, go to page 50.
HydroRanger 200 – INSTRUCTION MANUAL
7ML19985FC03