P650 Offset Calibration
Calibrates Empty (P006) if the reported level is consistently high or low by a fixed amount.
Primary Index
Values
Related
Before using this feature, verify the following parameters are correct:
•
Empty (P006)
•
Temperature (P664)
•
Offset Reading (P062)
•
Zero Head Offset (P605), if using OCM
Offset Calibration
Begin with a steady level.
1.
Press TRANSDUCER
2.
Repeat Step One at least five times to verify repeatability.
3.
Measure the actual reading (use tape measure).
4.
Enter the actual value.
The deviation between the entered Empty (P006) value and the calibrated Empty value is
stored in Offset Correction (P652).
P651 Sound Velocity Calibration
Changes the speed of sound constant.
Primary Index
Values
Related
Condition for use of this feature
•
The acoustic beam atmosphere is other than air
•
The acoustic beam atmosphere temperature is unknown
•
The Reading accuracy is acceptable at higher material levels only
For best results, calibrate with the level at a known value near empty.
7ML19985FC03
Single Point Model
Global
Range: -999 to 9999
•
P006 Empty
•
P062 Offset Reading
•
P605 Zero Head
to display the calculated reading.
Single Point Model
Global
Range: -999 to 9999
•
P653 Velocity
•
P654 Velocity at 20°C
HydroRanger 200 – INSTRUCTION MANUAL
Dual Point Model
Transducer
•
P652 Offset Correction
•
P664 Temperature
Dual Point Model
Transducer
Page 179