Measure:xmin - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

:MEASure:XMIN

Command Syntax
:MEASure:XMIN [<source>]
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | WMEMory<r>}
<n> ::= 1-2 or 1-4 (# of analog channels) in NR1 format
<m> ::= 1-4 in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:XMIN command installs a screen measurement and starts
an X- at- Min- Y measurement on the selected window. If the optional source
parameter is specified, the current source is modified.
:MEASure:XMIN is an alias for :MEASure:TMIN.
N O T E
Query Syntax
:MEASure:XMIN? [<source>]
The :MEASure:XMIN? query measures and returns the horizontal axis
value at which the minimum vertical value occurs. If the optional source is
specified, the current source is modified. If all channels are off, the query
returns 9.9E+37.
Return Format
<value><NL>
<value> ::= horizontal value of the minimum in NR3 format
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
"Introduction to :MEASure Commands"
":MEASure:XMAX"
on page 482
":MEASure:TMIN"
on page 1141
:MEASure Commands
on page 420
21