7
:ACQuire Commands
:ACQuire:POINts
Query Syntax
:ACQuire:POINts?
The :ACQuire:POINts? query returns the number of data points that the
hardware will acquire from the input signal. The number of points
acquired is not directly controllable. To set the number of points to be
transferred from the oscilloscope, use the command :WAVeform:POINts. The
:WAVeform:POINts? query will return the number of points available to be
transferred from the oscilloscope.
Return Format
<points_argument><NL>
<points_argument> ::= an integer in NR1 format
See Also
232
(see
page
1208)
"Introduction to :ACQuire Commands"
":DIGitize"
on page 199
":WAVeform:POINts"
on page 1032