:ACQuire:SEGMented:COUNt
Command Syntax
:ACQuire:SEGMented:COUNt <count>
<count> ::= an integer from 2 to 1000 (w/4M memory) in NR1 format
This command is available when the segmented memory option (Option SGM) is enabled.
N O T E
The :ACQuire:SEGMented:COUNt command sets the number of memory
segments to acquire.
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command, and data is acquired using the :DIGitize,
:SINGle, or :RUN commands. The number of memory segments in the
current acquisition is returned by the :WAVeform:SEGMented:COUNt?
query.
The maximum number of segments may be limited by the memory depth
of your oscilloscope. For example, an oscilloscope with 1M memory allows
a maximum of 250 segments.
Query Syntax
:ACQuire:SEGMented:COUNt?
The :ACQuire:SEGMented:COUNt? query returns the current count setting.
Return Format
<count><NL>
<count> ::= an integer from 2 to 1000 (w/4M memory) in NR1 format
See Also
Example Code
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
":ACQuire:MODE"
on page 231
":DIGitize"
on page 199
":SINGle"
on page 222
":RUN"
on page 220
":WAVeform:SEGMented:COUNt"
":ACQuire:SEGMented:ANALyze"
"Introduction to :ACQuire Commands"
"Example Code"
on page 236
on page 1039
on page 234
on page 227