Sbus<N>:Flexray:trigger:frame:ccrepetition - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

:SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition
Command Syntax
:SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition <cycle_count_repetition>
<cycle_count_repetition> ::= {ALL | <rep #>}
<rep #> ::= integer values 2, 4, 8, 16, 32, or 64
The :SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition command sets the
repetition number of the FlexRay cycle count (in the frame header) to
trigger on. The cycle count repetition setting is only valid when the
FlexRay trigger mode is set to FRAME.
Query Syntax
:SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition?
The :SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition? query returns the
current cycle count repetition setting for the FlexRay frame trigger setup.
Return Format
<cycle_count_repetition><NL>
<cycle_count_repetition> ::= {ALL | <rep #>}
<rep #> ::= integer values 2, 4, 8, 16, 32, or 64
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
"Introduction to :TRIGger Commands"
":SBUS<n>:FLEXray:TRIGger"
on page 935
on page 695
28
:SBUS<n> Commands
701