:CALibrate:STARt

Command Syntax
:CALibrate:STARt
The CALibrate:STARt command starts the user calibration procedure.
Before starting the user calibration procedure, you must set the rear panel CALIBRATION
N O T E
switch to UNPROTECTED, and you must connect BNC cables from the TRIG OUT connector
to the analog channel inputs. See the User's Guide for details.
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
"Introduction to :CALibrate Commands"
":CALibrate:PROTected"
on page 254
on page 258
:CALibrate Commands
259
9