Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual page 199

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

:DIGitize
Command Syntax
:DIGitize [<source>[,..,<source>]]
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | SBUS{1 | 2}}
<source> ::= {CHANnel<n> | DIGital<d> | POD{1 | 2}
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
<m> ::= 1-4 in NR1 format
The <source> parameter may be repeated up to 5 times.
The :DIGitize command is a specialized RUN command. It causes the
instrument to acquire waveforms according to the settings of the :ACQuire
commands subsystem. When the acquisition is complete, the instrument is
stopped.
If no argument is given, :DIGitize acquires the channels currently
displayed. If no channels are displayed, all channels are acquired.
The :DIGitize command is only executed when the :TIMebase:MODE is MAIN or WINDow.
N O T E
To halt a :DIGitize in progress, use the device clear command.
N O T E
MATH<m> is an alias for FUNCtion<m>.
N O T E
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
for the DSO models
| BUS{1 | 2} | FUNCtion<m> | MATH<m> | SBUS{1 | 2}}
for the MSO models
"Introduction to Root (:) Commands"
":RUN"
on page 220
":SINGle"
on page 222
":STOP"
on page 224
":TIMebase:MODE"
on page 925
Chapter
7, ":ACQuire Commands," starting on page 227
Chapter
33, ":WAVeform Commands," starting on page 1019
Root (:) Commands
on page 190
6
199