Acquire:type <Type; Acquire:count - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

7
:ACQuire Commands
:ACQuire:TYPE
Command Syntax
:ACQuire:TYPE <type>
<type> ::= {NORMal | AVERage | HRESolution | PEAK}
The :ACQuire:TYPE command selects the type of data acquisition that is to
take place. The acquisition types are:
• NORMal — sets the oscilloscope in the normal mode.
• AVERage — sets the oscilloscope in the averaging mode. You can set the
• HRESolution — sets the oscilloscope in the high- resolution mode (also
• PEAK — sets the oscilloscope in the peak detect mode. In this mode,
The AVERage and HRESolution types can give you extra bits of vertical
resolution. See the User's Guide for an explanation. When getting
waveform data acquired using the AVERage and HRESolution types, be
sure to use the WORD or ASCii waveform data formats to get the extra
bits of vertical resolution.
The obsolete command ACQuire:TYPE:REALtime is functionally equivalent to sending
N O T E
ACQuire:MODE RTIME; TYPE NORMal.
Query Syntax
:ACQuire:TYPE?
The :ACQuire:TYPE? query returns the current acquisition type.
Return Format
<acq_type><NL>
240
(see
page
1208)
count by sending the :ACQuire:COUNt command followed by the number
of averages. In this mode, the value for averages is an integer from 1 to
65536. The COUNt value determines the number of averages that must
be acquired.
The AVERage type is not available when in segmented memory mode
(:ACQuire:MODE SEGMented).
known as smoothing). This mode is used to reduce noise at slower
sweep speeds where the digitizer samples faster than needed to fill
memory for the displayed time range.
For example, if the digitizer samples at 200 MSa/s, but the effective
sample rate is 1 MSa/s (because of a slower sweep speed), only 1 out of
every 200 samples needs to be stored. Instead of storing one sample
(and throwing others away), the 200 samples are averaged together to
provide the value for one display point. The slower the sweep speed,
the greater the number of samples that are averaged together for each
display point.
:ACQuire:COUNt has no meaning.
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide

Advertisement

Table of Contents
loading

Table of Contents