Digital<D>:Position - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

:DIGital<d>:POSition
Command Syntax
:DIGital<d>:POSition <position>
<d> ::= 0 to (# digital channels - 1) in NR1 format
<position> ::= integer in NR1 format.
Channel Size
Large
Medium
Small
The :DIGital<d>:POSition command sets the position of the specified
channel. Note that bottom positions might not be valid depending on
whether digital buses, serial decode waveforms, or the zoomed time base
are displayed.
This command is only valid for the MSO models.
N O T E
Query Syntax
:DIGital<d>:POSition?
The :DIGital<d>:POSition? query returns the position of the specified
channel.
If the returned value is "- 1", this indicates there is no space to display the
digital waveform (for example, when all serial lanes, digital buses, and the
zoomed time base are displayed).
Return Format
<position><NL>
<position> ::= integer in NR1 format.
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
Position
0-7
0-15
0-31
"Introduction to :DIGital<d> Commands"
:DIGital<d> Commands
Top
Bottom
7
0
15
0
31
0