10
:CHANnel<n> Commands
:CHANnel<n>:INVert
Command Syntax
:CHANnel<n>:INVert <invert value>
<invert value> ::= {{1 | ON} | {0 | OFF}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:INVert command selects whether or not to invert the
input signal for the specified channel. The inversion may be 1
(ON/inverted) or 0 (OFF/not inverted).
Query Syntax
:CHANnel<n>:INVert?
The :CHANnel<n>:INVert? query returns the current state of the channel
inversion.
Return Format
<invert value><NL>
<invert value> ::= {0 | 1}
See Also
270
(see
page
1208)
"Introduction to :CHANnel<n> Commands"