12.3.2 Gas Delivery Schematic

Notes
M1046983 04/08
The Softkeys at the bottom of the Gas Delivery schematic bring up related diagnostic
screens that are also accessible from the Gas Delivery Subsystem menu.
• Gas Supplies — (Section 12.8.1)
• Mixer Output — (Section 12.8.2)
• Mixer Press/Temp — (Section 12.8.5)
Settings
Flow/Concentration
O2 Concentration
(if O2 Concentration selected)
Total Flow
(if O2 Concentration selected)
* O2 Flow
(if Individual Flow selected)
* Balance Gas Flow
(if Individual Flow selected)
* Total Fresh Gas Flow limited to 15.00 l/min
Remarks/Values
Move slider to increase/decrease value. Use arrow keys to
fine tune setting.
0%, 21%, 25% to 100%.
21% minimum for Air Balance Gas
0% minimum for N2O Balance Gas
Off, 0.10 l/min to 0.50 l/min (in 10 ml/min increments)
0.50 l/min to 15.0 l/min (in 0.25 l/min increments)
0.00 l/min to 15.00 l/min
Air = 0.00 l/min to 15.00 l/min
N2O = 0.00 l/min to 12.00 l/min