GE Datex-Ohmeda Aisys Carestation Reference Manual page 274

Anesthesia machine
Table of Contents

Advertisement

Aisys
7.10.12
Electronic
vaporizer 10VA
power
interconnect fault
isolation
7-92
Troubleshooting of the following Error Log Entries may lead you to further troubleshoot
the problem as detailed below:
• ADB 10VA POWER ERROR
• INFLOW ZERO 10VA POWER ERROR
• OUTFLOW SCAV 10VA POWER ERROR
• PROP VALVE HTR 10VA POWER ERROR
10VA power interconnect fault isolation procedure:
Continue to disconnect harnesses in the following order and retest until problem
resolved:
1. Disconnect the ADB harness at PCB.
• If problem goes away, replace ADB-PCB harness.
2. Disconnect the ACB harness at PCB.
• If problem goes away, replace PCB.
3. Disconnect the ACB harness at ACB.
• If problem goes away, replace ACB-PCB harness.
4. If cable disconnects do not eliminate the failure, replace ACB and retest.
04/08 M1046983