EVISION
8-6-3
Gas Spring Assy
Purpose: This is a description on how to remove and replace the Up/Down Assy.
8-6-3-1
Tools
•
Common pillips screwdrivers
•
Allen/Unbraco wrench
•
E-ring gripper
8-6-3-2
Needed Manpower
•
1 person, 45 minutes + travel
8-6-3-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 3.
•
Removal of following parts should be proceeded before start.
•
Remove CRT Assy, KeyBoard Assy, Probe Holder,Left Cover, Right Cover, Front Base Cover,
Front Cover, OP Side L Cover, OP Side R Cover, OP Rear Cover, EMI Cover L, EMI Cover R,
BackPlane, and BEP (Back End Processor) ASSY.
8-6-3-4
Removal Procedure
1.) Lower the OP Frame down to its lowest position. Refer to Figure 8-80.
2.) Unscrew three (3) screws (1-3) to remove the OP push button.
2
Release Button
Figure 8-80 Lowering the OP Frame
Chapter 8 Replacement Procedures
PRESS
LOGIQ™5 S
M
ERVICE
ANUAL
8-79