Test 12
Loop
M1046983 04/08
Testing the inspiratory side of the circuit module
Note: If required, set up the machine as in Test 10 and 11.
1. Connect the Circuit Test Tool to the Circuit Module as shown above.
2. Insert an appropriate test plug in the inspiratory outlet to the canister as shown
above.
3. Set O
Flow to 0.2 l/min.
2
• Ensure that the Airway Pressure rises to ≥ 30 cm H
4. Set O
Flow to 0.00 l/min.
2
7 Troubleshooting
Plug
O.
2
7-25