5.1 Primary Regulators

w w w w WARNING
5-2
First, follow the procedure in Section 5.1.1 to gain access to the regulators.
Then, in Section 5.1.2, select the test that is appropriate for the regulator you
are testing.
When testing/adjusting N
system. Use a safe and approved procedure to collect and remove it.
Note
To test or calibrate the primary regulators, you must set the system to the
Install/Service mode and use the PC based Service Application to control
flow through the regulator.
O regulators, nitrous oxide flows through the
2
04/08 M1046983