12.3.4 Vaporizer Schematic

Notes
M1046983 04/08
The Softkey at the bottom of the Vaporizer schematic brings up a related diagnostic
screen that is also accessible from the Vaporizer Subsystem menu.
• Vaporizer Output — (Section 12.10.1)
Settings
Manual Control
Agent Concentration
or
Prop Flow Valve DAC Value
Zero Valves Auto
Inflow Zero
Outflow Zero
Remarks
Set "Agent Concentration" to "Off" to enable Manual
Control
Move slider to increase/decrease value. Use arrow keys to
fine tune setting.
Select "Zero Valves Auto" for best accuracy flow
measurements.
Not settable if "Zero Valves Auto" is checked.
Outflow Meter reading are compensated by ambient
pressure stored at machine start-up.