GE Datex-Ohmeda Aisys Carestation Reference Manual page 83

Anesthesia machine
Table of Contents

Advertisement

Mechanical inspiration
(EZchange and
condenser
ON)
D
P
AP Airway Pressure
D
Drive gas
FG Fresh Gas
P
Pilot pressure
V
Bag/Vent switch to Vent
1
Flow to absorber
2a
Flow from absorber
2b
Flow from condenser
3
Inspiratory flow
Figure 2-38 • Mechanical inspiration through Condenser with EZchange Canister and Condenser ON
M1046983 04/08
The Bag/Vent switch closes the manual path (V). Pilot pressure (P) closes the
exhalation valve.
Drive gas (D) pushes down on the bellows. Gas flows from the bellows (1), through the
absorber (2a), Condenser (2b), and through a unidirectional valve (inspiratory check
valve) to the patient (3).
During inspiration, fresh gas flows into the inspiratory limb, upstream of the inspiratory
check valve.
D
1
V
AP
FG
2b
2a
u
2 Theory of Operation
3
2-51