Evap Troubleshooting Flowchart - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

7.10.3 eVap Troubleshooting Flowchart

Save the
Test
Cassette
Results for
future
reference.
Download and combine the
Machine Logs or review the
logs in Service Mode.
Refer to Transfer Logs.
(Section 12.6.3)
Analyze
the logs.
No
Compare the
Combined log to the
Error Logs in the
"Electronic Vaporizer Table".
(Section 7.10.4)
Proceed to the appropriate
troubleshooting flowchart.
19
Flowchart
M1046983 04/08
Yes
Is the Error prefix
"VAPCHK"?
Start
Within the PC Service Application,
perform the Vaporizer Test.
(Section 12.10.2)
Does theVaporizer Test pass?
No
Yes
7 Troubleshooting
Follow the Corrective
Action in the Vaporizer
Test Results
troubleshooting procedures.
(Section 7.10.2)
Follow the Corrective
Action in the Vaporizer
Checkout Troubleshooting.
(Section 7.10.13)
7-83

Advertisement