2.6 Power Controller and Anesthesia Control board connections

Power Controller board
Anesthesia Control board
M1046983 04/08
The Power Controller:
• Distributes 12.5 VDC power and communicates with the Display Unit (by way of the
Display Connector board) through connector (10).
• Distributes 12.5 VDC power to the Anesthesia Control board through connector (14).
The Anesthesia Control board:
• Receives power from the Power Controller board through connector (14).
• Distributes 10VA power supplies to the Pan Connector board through connector (15).
• Communicates with Pan assemblies through connector (16).
• Communicates with Display Unit through connector (12).
• Distributes 10VA power supplies to the Display Unit through connector (11).
15
14
14
10
11
16
2 Theory of Operation
12
2-15