9.17 Replace ACGO selector switch

Removal
5
3
2
Replacement
M1046983 04/08
1. Remove the tabletop (Section 9.4).
2. Clip the tie wraps (
1
4
3. Disconnect the fresh gas (
4. Disconnect the wires from the ACGO mode microswitch (
switch.
5. Disconnect the wires from the flush pressure switch (
6. Set the ACGO selector switch to ABS.
7. Back out the selector switch mounting screws (
of the mounting casting.
8. While pushing the selector knob toward the machine and holding it steady, push the
valve body toward the knob and rotate it counterclockwise to separate the valve
body from the knob assembly.
9. Remove the knob assembly and protective shroud from the machine.
10. Remove the valve from the silicone output tubes.
1. Remove the knob assembly from the valve body.
2. Back out the selector switch mounting screws until the tips are flush with the face of
the mounting casting.
3. Guide the outlet fittings of the valve body into their respective silicone tubes.
4. Hold the selector knob with the
indicator mark facing down. Turn the
chrome collar to its maximum
counterclockwise position (as
viewed from the front).
) from the outlet barb fittings at the side of the switch.
1
6
) and flush (
) tubes at the back of the switch.
2
3
9 Repair Procedures
) at the back of the selector
4
) on top of the selector switch.
5
) until the tips are flush with the face
6
9-45