Aladin Cassette Parts Replacement - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

9 Repair Procedures
9.8.2 Aladin
You can replace the handle assembly (
) and the o-ring (
) for the filler cap
Item 1
Item 2
2
without disassembly. To prevent scratching the cap sealing surface, use caution when
cassette parts
removing the o-ring.
replacement
1
2
For the remaining components, follow the outlined sequence to access the individual
components.
1. Ensure that the cassette is empty of agent (refer to Section 9.8.1).
2. Remove the handle assembly.
3. For non-DES cassettes, remove the filler cap.
4. Turn the cassette over and remove the bottom plate (
).
Item 3
• Note that the DES cassette does not use a bottom plate.
3
5. Remove the label (
) to access the mounting screw (
) for the mask.
Item 4
Item 5
5
4
M1046983 04/08
9-17