Aisys
Error Log Entry
Action/Troubleshooting
BATTERY CHARGE FAIL
Check the battery charge circuit in Service Software.
Replace Battery.
BATTERY EMPTY
Leave the system plugged in to charge the battery.
If problem continues, check the battery charge circuit in Service Software.
Replace Battery.
BATTERY FAIL
Leave the system plugged in to charge the battery.
If problem continues, check the battery charge circuit in Service Software.
Replace Battery.
BATTERY LOW
Leave the system plugged in to charge the battery.
If problem continues, check the battery charge circuit in Service Software.
Replace Battery.
BATTERY MISSING
Connect Battery.
Leave the system plugged in to charge the battery.
If problem continues, check the battery charge circuit in Service Software.
Replace Battery.
BATTERY REVERSED
CONNECTIONS
Check Battery Connections.
7-50
Alarm Text
Condition (Basic info)
No battery backup
The system is in standby and
the battery charge current is
>4.0 amps.
or
The system is powered on
with a battery current >1.3
amps.
System shutdown in
Available battery power is
<5 min
between 1 and 5 minutes
No Battery Backup.
Battery voltage <10.5 V
or
While in bulk, over, or float
charging battery is <10.5VDC
or
Battery has been bulk
charging for >12 h in Standby
or 24 h while powered on.
or
Voltage > 16.5V during bulk
or over charging and normal
current >0.25 Amps
Plug in power cable.
Available battery power
On battery
decreases to between 20 and
30 min
No battery backup
Any battery voltage is
between ±1.0 VDC.
No battery backup
Any battery voltage is less
than –1.0 VDC
Priority Source Enabling Criteria
Medium
PC
High
PC
AC Mains Power Failure
in progress.
Medium
PC
Medium
PC
Mains AC Mains Power
Failure in progress.
Medium
PC
POST state
Medium
PC
04/08 M1046983