Remove The Electronic Vaporizer - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

Aisys
9.7.1 Remove the
electronic
vaporizer
w
CAUTION
9-12
1. If the eVap is equipped with bulkhead fittings on the eVap support bracket,
disconnect the inlet, outlet, and scavenging tubing from the bulkhead fittings.
Remove the suction tubing connectors (if equipped) from the clips on the bottom of
the eVap support bracket.
Suction Tubing Clips
If the eVap is not equipped with bulkhead fittings, disconnect the tubing directly from
the inlet, outlet, and the scavenging ports of the eVap.
2. Disconnect the pan connector harness (
3. Clip the tie-wrap holding the pan connector harness to the eVap support bracket.
4. Remove the six screws (
(B)
(B)
5. Note: The eVap assembly weighs approximately 5 kg (11 lb). Most of the weight of
the eVap is at the rear.
When handling the eVap, be careful not to damage the Flow Control Valve at the
back of the flowmeter block.
6. To remove the eVap:
• Firmly take hold of the eVap at the rear with one hand without touching any
boards.
• Support the front of the eVap with the other hand.
• Raise the rear of the eVap tilting the dashboard forward.
• When clear, remove the eVap from the machine.
) from the Agent Delivery board.
A
) holding the eVap to the dashboard.
B
(A)
Scavenging
Outlet
Inlet
04/08 M1046983