4.5.4 Insp Flow Zero

Instructions
Insp Flow Zero menu
4-22
Note: Not required for machines with System Software 4.X or greater.
The Insp Flow Zero instructions appear when the focus is on the Insp Flow Zero
menu item.
Refer to Section 5.4.4, "Insp Flow Zero."
Read all steps before you start:
1. Push the ComWheel to start the zero check
2. No disassembly is required.
• If the outcome of the calibration is Pass, the new calibration data is saved.
• If the outcome is Fail, the old calibration data is retained.
• The result of the calibration is saved to the Event Log.
Selecting Previous Menu before the calibration is done aborts the calibration
in progress and keeps the old calibration constants.
Menu Item
Message text
Start
If the result is failed, do the insp flow
valve calibration.
Previous Menu
Return to previous menu.
During Calibration:
"Calibration in progress. Push
ComWheel to cancel."
Values
Pass or Fail