GE Datex-Ohmeda Aisys Carestation Reference Manual page 480

Anesthesia machine
Table of Contents

Advertisement

Aisys
12.7.2 Power Diagnostics
Anesthesia Control Board Power
Label
Value
12 Vdc Supply
XX.XX
ADC Ref
X.XXX
Label
Gas Select 10VA Volts
Gas Select 10VA Amps
Press Transducer 10VA Amps
Vent Interface Board 10VA Volts
Vent Interface Board 10VA Amps
Vent Valves 10VA Volts
Vent Valves 10VA Amps
Acces 1 10VA Volts
Acces 1 10VA Amps
Gas Unit 10VA Volts
Gas Unit 10VA Amps
Mixer 10VA Volts
Mixer 10VA Amps
Alt O2 10VA Volts
Alt O2 10VA Amps
12-14
Anesthesia Control Board Power
Units
Normal range
Vdc
11.90 to 12.90
Vdc
4.018 to 4.176
Value
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
Label
Periph1 10VA Volts
Periph1 10VA Amps
Periph2 10VA Volts
Periph2 10VA Amps
Agent Delivery 10VA Volts
Agent Delivery 10VA Amps
Agent Deliv Prop Valve 10VA Volts
Agent Deliv Prop Valve 10VA Amps
Agent Deliv Inflow and Zero 10VA Volts
Agent Deliv Inflow and Zero 10VA Amps
Agent Deliv Outflow and Scav 10VA Volts
Agent Deliv Outflow and Scav 10VA Amps
Mini AA 10VA Volts
Mini AA 10VA Amps
Bellows Position Sensor 10VA Volts
Bellows Position Sensor 10VA Amps
Value
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
04/08 M1046983