9 Repair Procedures
9.12.4 Backup
To remove the batteries
batteries
1. Loosen the two screws (circled) that hold the battery retainer to the enclosure.
2. Remove the retainer.
3. Disconnect the flex-cable from the Power Controller board (A).
A
4. Remove both batteries simultaneously as a unit from the machine.
w
CAUTION
To avoid damaging the flex-cable, handle it with care. Support both batteries to
avoid tearing the flex-cable.
To replace the batteries
1. Transfer the flex-cable to the new batteries.
2. To replace the batteries, reassemble in reverse order
3. Allow the batteries to charge.
4. Recycle old batteries in same packaging according to local requirements.
M1046983 04/08
9-35