Vaporizer Subsystem Menu - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

Aisys

12.10 Vaporizer Subsystem menu

12.10.1 Vaporizer Subsystem
12-28
The Vaporizer menu includes the following menu items:
• Vaporizer Schematic (Section 12.3.4)
• Vaporizer Output (Section 12.10.1)
• Perform Vaporizer Test (Section 12.10.2)
Vaporizer Output
Vaporizer Output
Label
Cassette Temp 1
Cassette Temp 2
Manifold Temp 1
Manifold Temp 2
Input Flow
Output Flow
Valve Drive Sense
Cassette Pressure
Agent ID
Agent %
Agent Level
Value
Units
XX.X
C
F
XX.X
C
F
XX.X
C
F
XX.X
C
F
XX.X
ml/min
XX.X
ml/min
XX.X
mA
XX
kPa
psi
bar
None, ENF,
DES, HAL, ISO,
SEV, TEST,
INVALID
XXX
Full,
3/4 Full*,
1/2 Full,
1/4 Full,
Empty,
Overfilled,
Not Available
* System software 3.X only
Range
-2.5 to 50.0
27.5to 122.0
-2.5 to 50.0
27.5to 122.0
2.5 to 50.0
27.5to 122.0
2.5 to 50.0
27.5to 122.0
-200 to 6500
-200 to 6500
N/A
77.89 to 192.51
11.30 to 27.92
0.78 to 1.93
ENF, ISO, HAL
• 0.0%, 0.2% to 5%
SEV
• 0.0%, 0.2% to 8%
DES
• 0.0%, 1% to 18%
04/08 M1046983