Gas Spring Replacement Procedure

8-6-9-1

Tools

Use the appropriate Phillips screwdriver, light hammer, punch, and 5 mm Allen key.
FRU Part #
Refer to
8-6-9-2
Time Required
120 min
8-6-9-3
Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-6-9-4
Gas Spring Removal Procedure
1) Perform the following removal procedures:
2) Unscrew the shaft covering screws located on either side of the main column - see
8-160
Section 8-6 - Mechanical Platform Components - Replacement Procedures
1
EVISION
Table 9-6
.
on page 9-5
21.5" Monitor Replacement Procedure on page 8 - 31
Operator Panel Keyboard Assembly Removal Procedure on page 8 - 44
OPIO Basket Removal Procedure on page 8 - 26
Figure 8-194 Shaft Covering Screws
V
S60N/V
S70N
IVID
IVID
Power Shut Down on page 4 -
(if attached to the system)
B
S
M
ASIC
ERVICE
ANUAL
7.
Figure
8-194. .