4.2 Install/Service Menu (Super User)

Install/Service - Page 1
Menu Item
Colors and Units
Cumulative Gas Usage
Show Alarm Limits
External Gas Monitor
Save Default Case
Factory Defaults
Parameter Settings
Install/Service - Page 2
M1046983 04/08
Use the super-user password to access the Install/Service menu:
"16-4-34."
Message text
Trends Setup
Configure graphical trend pages.
Set colors and units of parameters.
View total usage of fresh gases since
last usage reset.
Select yes to show alarm limits in
digit fields.
Yes disables O2 limit alarms, the "No
O2 sensor alarm", and the "No CO2
or AA monitor" alarm.
Save normal screen, gas settings,
circuit type, ventilator, and alarm
settings from the last case or
presettings as defaults
Return to default factory settings.
After selecting Factory Defaults:
"Reset machine for defaults to take
effect."
Set volume conditions and CO2
humidity compensation.
Show page 2 of the Install/Service
Menu.
Exit
Turn power off to exit the
Install/Service menu.
4 Install/Service Menus
Comments
Refer to section 4.2.1
Refer to section 4.2.3
Default is Yes.
Default is No.
Select yes only if sys is using external
monitor for O2, AA, and CO2.
Last used alarm settings (including hide/
show alarm limits, Auto MV Limit, alarm
volume), screen layout (middle waveform
selection, sweep speed), ventilator mode
and setting, balance gas, and start case
gas outlet selection are saved as facility
defaults.
Note: The Pmax alarm limit shall not be
saved higher than 40 cmH2O. The low
FiO2 alarm limit shall not be saved lower
than 21%.
Action: All facility defaults get replaced
with factory defaults. Super User settings
also get set to Factory Defaults. No
Service level configuration settings are
changed.
4-3