Test 14
Note: To ensure a air-tight
seal, use the corresponding
plug as illustrated for the
original flow sensor ( A )
or the new,
moisture resistant
(offset) flow sensor ( B ).
M1046983 04/08
Testing the flow sensors only
1. Remove the Flow Sensor Module.
2. Plug each Flow Sensor as shown above.
3. Connect the low-pressure leak test device to the open end of the Flow Sensor.
4. Block the connector end of the Flow Sensor with your hand.
5. Compress and release the bulb until it is empty.
6. If the bulb inflates in 30 seconds or less, there is a leak in the flow sensor.
7. If there are no leaks in the flow sensors, go to Test 15.
A
B
7 Troubleshooting
7-27