GE Datex-Ohmeda Aisys Carestation Reference Manual page 206

Anesthesia machine
Table of Contents

Advertisement

Aisys
Test 10
Test 11
7-24
Testing the circuit module and the canister
1. Remove the Flow Sensor module.
2. Connect the Circuit Test Tool to the Circuit Module as shown above.
3. Set O
Flow to 0.2 l/min.
2
• Ensure that the Airway Pressure rises to ≥ 30 cm H
4. Set O
Flow to 0.00 l/min.
2
Testing the circuit module
Plug
Note: If required, set up the machine as in Test 10.
1. Remove the Soda Lime Canister.
2. Using appropriate Test Plugs, plug the three canister ports in the Circuit Module as
shown above.
3. Set O
Flow to 0.2 l/min.
2
• Ensure that the Airway Pressure rises to ≥ 30 cm H
4. Set O
Flow to 0.00 l/min.
2
O.
2
Plug
Plug
O.
2
04/08 M1046983