GE Datex-Ohmeda Aisys Carestation Reference Manual page 218

Anesthesia machine
Table of Contents

Advertisement

Aisys
ACB - Mixer Troubleshooting
See "Power /
Valves / Mixer
Troubleshooting"
(Flowchart 16)
Contact
Technical
Support
Problem Solved
XMT
Check / Replace Cables.
Replace ACB if problem persists.
13
Flowchart
7-36
Start
Does Mixer communicate
in Service Application?
No
Yes
Can the issue be
duplicated?
No
Yes
Check / Replace
ACB to Pan
Connector and
Pan Connector to
Mixer
Problem Continues?
No
Yes
Check the Mixer
XMT / RCV LEDs
Does either LED indicate
MIXER ACB COM FAIL
MIXER ACB COM FAILLOST CMD
Activity?
RCV or No Activity
Replace Mixer
Error Entry
04/08 M1046983