Error Log Entry
Action/Troubleshooting
Circuit check failed.
Perform suggested action and repeat the system Checkout.
Circuit O2 check
skipped.
Perform suggested action and repeat the system Checkout.
COM ERROR VENT TO
ACB
Reboot System. If problem continues:
1. Check cabling.
2. Replace VIB.
3. Replace ACB.
Compatibility failure:
No version info in file
for subsystem 0.
Look for other entries in the Error Logs. i.e. "Self-tests Failed".
Perform Software Download.
Compatibility failure:
Software Error
Look for other entries in the Error Logs. i.e. "Self-tests Failed".
Perform Software Download.
M1046983 04/08
Alarm Text
Condition (Basic info)
A message or failure
displayed during the system
checkout.
A message or failure
displayed during the system
checkout.
System Malfunction
After regular communications
has been established
between the Ventilator
boundary object and the Vent
SIB CPU, a total loss of
communications shall be
declared if the Ventilator
boundary object receives no
messages from the Vent SIB
CPU for 35 milliseconds.
Indicates a subsystem did
not report compatibility
information to the Display
Unit.
Indicates a subsystem did
not report compatibility
information to the Display
Unit.
7 Troubleshooting
Priority Source Enabling Criteria
DU
DU
High
AC Vent
DU
DU
7-53